What does the instrument documentation actually say?

NI documents a specific behavior for PXIe-4081 waveform acquisition: an input outside the measurement range produces invalid NaN values, and recovery can include further invalid points. Its warning is therefore part of the evidence, not a numerical measurement of how far the signal went beyond the range.

Commissioning Docs is independent and is not affiliated with NI. Other instruments and export formats can behave differently. Read the actual model, function and software documentation rather than apply this example to every meter or historian.

Sources: NI: NI PXIe-4081 documentation: Acquiring Current and Voltage Waveforms, checked 20 October 2026

Which states should the result register keep separate?

Keep the original export and its status fields. If a spreadsheet cannot represent the instrument status, retain a separate linked status column instead of replacing invalid data with a convenient number.

Which states should the result register keep separate?
Result stateEvidence to keepReview disposition
Valid readingValue, unit, time and statusAssess against the approved criterion
Overrange warningExact warning and channelPeak remains unresolved
Invalid sampleOriginal marker and export definitionDo not substitute zero
Recovery intervalActual instrument guidance and dataDo not invent a universal duration
ReacquisitionApproved new test and result referenceKeep the original exception
AcceptanceAuthorized technical decisionNo automatic pass or fail from a blank

How can a summary conceal the problem?

Illustrative example. A fictional export contains valid values of 8 and 9 signal units, with an invalid overrange entry between them. A spreadsheet replaces the invalid entry with zero and reports 9 as the maximum for the entire interval. Nine is the largest retained valid number; the evidence does not establish the true peak during the invalid interval.

The reviewer retains all three original states, labels the interval incomplete and asks the commissioning lead for an approved disposition. A larger graph axis, interpolation or copied neighbor value does not recover the missing measurement. The example provides no equipment limit or real measured result.

What should happen before the record is closed?

Identify which requirement needs a valid peak, how the warning affects that requirement and who can approve further work. Any repeat measurement needs an approved method, appropriately selected equipment and the site safety process. Do not change ranges on energized equipment from a generic web instruction.

This is an evidence-control worksheet, not an instrument selection or test procedure. Engineering sign-off is performed only by the registered engineer on the job.

Common questions

What else should the bid team know?

Is an invalid overrange entry a measured zero?

No. Preserve its actual status and meaning.

Can the true peak be reconstructed from the range label alone?

No. The label is not the missing waveform.

Does this article prescribe a recovery time?

No. Obtain the actual instrument guidance and technical disposition.

Primary sources

  1. NI PXIe-4081 documentation: Acquiring Current and Voltage Waveforms, checked 20 October 2026NI